<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>17723</ErrorNumber>
<ErrorFamily>Exposure Head</ErrorFamily>
<ErrorMessage>
HEAD: Command parameter error
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Thermal</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
A command was sent to the head that specified an incorrect parameter.
This could be due to a incorrect nvs parameter, but in reality, this
is a firmware bug. Please try to duplicate the error and send the
diagnostic log to the firmware engineer.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Call service.
</text></DescriptionText>
</OperatorAction>
<Cause>
Possible incorrect nvs parameter setting.
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:44 2008 PST
</Updated>
<Links>
</Links>
</Error>